Sensor Subsystem Tests

Salinity Test

Objective:

Achieve a salinity sensor accuracy within -/+ 5 ppt using Vernier Salinity Sensor in a controlled environment. 

 

Procedure:

1. Carefully make various salinity solutions from 0 - 35 (ppt) with distilled water and NaCl 

2. Conduct two-point calibration with LabQuest 2 (manufacturer's calibration equipment)

3. Test accuracy by connecting salinity sensor to MCU and reading the known salinity solutions

Set-up Circuit Schematic:

sali.png

Salinity Test Results

salinity_solutions.png

Salinity Solutions

salinity_mean.png
35ppt.png

Our Vernier Salinity Sensor had the most discrepancies when measuring a known solution of 35 ppt. The error was -/+ 3.567 ppt and ultimately Still less than -/+ 5 ppt

SONAR Test

Objective:

Measure the Percent (%) error of SONAR sensor in 

pool at different depths. We want to achieve a % error less than 5%. 

 

Procedure:

1. Measure the known depth with tape ruler.

2. Connect sonar sensor to MCU Rx/Tx.

3. Upload code and open serial monitor. 

4. Submerge sonar sensor into pool. 

5. Read serial monitor only at 100% confidence.

Set-up Circuit Schematic:

sonar.png

SONAR Test Results

sonar_test.png

As showcased in the graph, our highest %error value is 4.1%. Therefore the Test has passed. (4.1<5)